Peak
SCPI Command :
FETCh:LTE:MEASurement<Instance>:MEValuation:DMARker<No>:EVMagnitude:PEAK
- class PeakCls[source]
Peak commands group definition. 1 total commands, 0 Subgroups, 1 group commands
- fetch(xvalue: int, trace_select: TraceSelect, deltaMarker=DeltaMarker.Default) float [source]
# SCPI: FETCh:LTE:MEASurement<Instance>:MEValuation:DMARker<No>:EVMagnitude:PEAK value: float = driver.multiEval.dmarker.evMagnitude.peak.fetch(xvalue = 1, trace_select = enums.TraceSelect.AVERage, deltaMarker = repcap.DeltaMarker.Default)
Uses the markers 1 and 2 with relative values on the bar graphs: EVM RMS, EVM peak, magnitude error and phase error vs SC-FDMA symbol.
Use RsCmwLteMeas.reliability.last_value to read the updated reliability indicator.
- param xvalue:
(integer or boolean) integer X value of the marker position relative to the X value of the reference marker There are two X values per SC-FDMA symbol on the X axis (symbol 0 low, symbol 0 high, …, symbol 6 low, symbol 6 high) . Range: -13 to 13
- param trace_select:
CURRent | AVERage | MAXimum
- param deltaMarker:
optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Dmarker’)
- return:
yvalue: float Y value of the marker position relative to the Y value of the reference marker